Skip to main content

This product is not available in your country

Back

Flue gas analyser EUROLYZER® S1 1./44. BImSchV set O2, CO/H2+, NO, draft | P00144

Flue gas analyser EUROLYZER® S1 1./44. BImSchV set

manufactured goods

  • TÜV-tested measurement as per 1st and 44th BlmSchV (German Federal Immission Act)
  • automatische Dichtheitsprüfung von Messgerät und Sonde nach VDI 4207 Blatt 1
  • H2-compensated robust CO sensor (measuring range up to 10,000 ppm)
  • Data transmission via Bluetooth® or QR code to the AFRISO app EuroSoft® connect for on-site documentation
  • large high-resolution display for optimum readability and and intuitive menu navigation, protected by scratch-resistant glass
  • Pioneering "dual-mode concept": Intuitive one-hand operation via scroll wheel or touch display – sensitive even with work gloves
  • Ready for all typical HVAC measuring tasks thanks to Bluetooth® connection to more than 35 sensor modules and sets of the CAPBs® system
  • configurable macros for repetitive tasks
  • Multi-memory: Combine up to six different measurements into a single measurement report (for example, full load measurement/partial load measurement or flue gas measurement and flow pressure measurement)
  • 5 years warranty on O2 sensor
  • 7 years full warranty in connection with a service contract

Designed for flue gas measurements at oil, gas and pellets fired systems as well as for tightness tests at gas lines, for checking the connection pressure/flow pressure and the pressure loss. TÜV-tested measurements as per 1st and 44th BlmSchV (German Federal Immission Act) can be performed reliably and intuitively. Fast and easy Bluetooth® connection to more than 35 sensor modules and sets of the AFRISO CAPBs® make EUROLYZER® S1 the central controller for all HVAC tasks.

Zielgruppe
Einsatzmedien
Produkteigenschaften
Schnittstellen
Zulassungen

measuring ranges (measured values)
Flue gas temperature/differential temperature
0/1 150 °C
Air temperature/combustion air temperature
-20/+200 °C
Draft
-70/+70 hPa
O2 measurement
0/21 Vol.-%
COH2 measurement
nominal:
0/4 000 ppm
maximum:
0/10 000 ppm
NO measurement
0/5 000 ppm
additional technical specifications
Indication (calculated values)
CO2, COunverdünnt, Lambda, Taupunkt, Eta efficiency, Abgasverluste qA, CO mg/m3 (natural gas), CO mg/kWh (Heizöl)
Operating temperature range
Ambient:
0/40 °C
Storage:
-20/+50 °C
weight (measuring instrument)
  540 g
Dimensions
W x H x D:
80 x 218 x 44 mm
Degree of protection
IP 40 (EN 60529)
Display
high-resolution touch display 4.3 "
Connections
Draft:
∅ 7 mm
Gas:
∅ 8 mm
temperature:
2 x socket type K
Supply voltage
lithium-ion battery x (3.6 V/5000 mAh)
Power supply unit (USB-C)
Battery or mains
Hours of operation (eco mode)
  12 Stunden
Data memory
internal memory,   10000 Messungen
Interfaces
Infrared,
Bluetooth®,
QR code generator,
USB-C
Approvals
TÜV-tested as per 1st and 44th BImSchV,
DIN EN 50379-1/2
Scope of delivery
  • Flue gas analyser EUROLYZER® S1 (O2, CO/H2+, NOx, draft, pressure)
  • mit integrierten Haftmagneten im Gehäuse
  • Calibration report
  • Schnelllade-Netzteil NTE 5 SF
  • charging cable USB-A to USB-C (1 m)
  • 5 x Infiltec-Feinfilter
  • 5 x Teflon membrane
  • Kondensatfilterpatrone KFP
  • Base handle AWS-B for interchangeable probes
  • Wechselsonde AWS-S 300 mm
  • Interchangeable probe AWS-S 750 mm
  • Verbrennungsluftfühler TFB-VL 600-1
Optional auf Anfrage
  • CO-Spülpumpe
  • automatische Dichtheitsprüfung von Messgerät und Sonde nach VDI 4207 Blatt 1
  • Messzellen: COhigh, NO2, SO2

SDB_525226-168-1013.PDF

SDB_525226-168-1013.PDF

PDF
Download
UN38-3_525226-168-1013.PDF

UN38-3_525226-168-1013.PDF

PDF
Download
MIETVERTRAG_MESSGERAETE_01-23.PDF

rental contract handheld measuring instruments EUROLYZER, MULTILYZER, STM 225, DPK 60-7 as well as EUROprinter

PDF
Download
WARTUNGSVERTRAG_EUROLYZER_S1_01-23.PDF

Maintenance contract EUROLYZER® S1

PDF
Download
900_000_1087-01.PDF

Flue gas analyser EUROLYZER® S1

PDF
Download