Skip to main content
Back

RF100Ch MK1 D402 | 87618402

Rohrfeder-Chemiemanometer mit Grenzsignalgeber Typ D4

manufactured goods

Range0/40 bar
housing ⌀ (nominal size)100 mm
ConnectionG½B
alignmentbottom
accuracy class1,0
Part no.87618402

  • measuring system fully welded to housing
  • robust mechatronical pressure gauge
  • Up to four switching contacts
  • tightness-tested with helium
  • GOSSTANDART-certified

For corrosive gaseous and liquid media which are not highly viscous and do not crystallise; suitable for corrosive environments.

Type
D4
Nominal size
100 - 100
accuracy class (EN 837-1/6)
1,0
ranges (EN 837-1/5)
see ordering table
application area
static load
≤ 600 bar:
Full scale value
> 600 bar:
¾ x full scale value
dynamic load
≤ 600 bar:
0.9 x full scale value
> 600 bar:
⅔ x full scale value
short-term
≤ 600 bar:
1.3 x full scale value
> 600 bar:
Full scale value
contact types
Magnetic spring contact (MK)
minimum range
Contact
MK Single: 1.6 bar
Operating temperature range
Medium:
max. 150 °C
Ambient:
-20/+60 °C
Temperature performance
Indication error when the temperature of the measuring system deviates from the normal temperature of 20 °C:
bei Temperaturzunahme ca. ± 0,4 %/10 K,
falling temperature approx. ± 0.4 %/10 K
vom jeweiligen Skalenendwert
Degree of protection
IP 54 (EN 60529)

Standard design

Connection
stainless steel 316 Ti/316 L, bottom
G½B, SW22 (EN 837-1/7.3)
Electrical connection
Junction box
Measuring element
Bourdon tube, stainless steel 316 Ti/316 L
≤ 60 bar:
"C" type tube
> 60 bar:
helical tube
movement
Stainless steel
Dial
Aluminium, weiß
Scaling black
Pointer
Aluminium, schwarz
Housing
Edelstahl 304, with blow-out
bayonet bezel
Edelstahl 304
window
Kunststoff (Makrolon), with contact adjustment lock
Inductive contact (IK)
Magnetic spring contact (MK)
Supply voltage
max. AC/DC 250 V
Making current and breaking current
max. 1 A
Continuous current
max. 0.6 A
Switch rating
max. 50 VA, max. 30 W
Contact
Ag80 Ni20 Au 10 μm
Switching accuracy
approx. 2–5 % of full scale value
Operating temperature range
-20/+70 °C or corresponding to the respective measuring instruments
Adjustment range
5–95 % measuring range of gauge

Click the corresponding image for detail views, additional information and dimension tables.

854_017_0336-00.PDF

Rohrfeder-, Kapselfeder-, Plattenfeder-, Membranfedermanometer für Differenzdruck