Skip to main content
Back

BIOLYZER | 69643

Gasanalysegerät für diskontinuierliche Messung

manufactured goods

  • for discontinuous measurement
  • ideal for biogas plants
  • monitors up to four gas components
  • compact wall mounting housing system ready for installation

For discontinuous analysis and regular process monitoring of biogenous process gases such as biogas, sewage gas and landfill gas. All important gas types such as methane, hydrogen sulphide, oxygen and carbon dioxide can be monitored.

Gas measuring system for discontinuous, selective measurement and indication of up to four different gas components (CH4, H2S, O2 and CO2). CH4 and CO2 are detected by means of infrared technology, O2 and H2S by means of electrochemical sensors. All components are integrated in a robust wall mounting housing. BIOLYZER features an LED display, status indicators for each measurement channel as well as a lit LC display. The memory has a history function, the values can be displayed. With freely adjustable alarm thresholds and binary outputs for: active, error, calibration, alarm. The gas treatment system with all important components (gas cooler with hose pump, fine filter, aerosol filter, rotameter with needle valve, anti-detonation device) is integrated in a robust, air-flushed wall mounting housing. BIOLYZER is delivered complete with wiring, hoses, calibration and ready to be mounted.
BIOLYZER LT is suitable for simple routine checks. Version without gas cooler and load limitation for hydrogen sulphide measurements. Therefore, the standard measuring range is 0 to 1,000 ppm H2S.

gas types/measuring ranges
CH4:
0/100 Vol.-%
CH4:
IR double beam
CO2:
0/100 Vol.-%
CO2:
IR double beam
O2:
0/25 Vol.-%
O2:
electrochemical
H2S
BIOLYZER: 0/5 000 ppm
H2S:
electrochemical
Measuring intervals can be programmed for the individual gas types. Manual measurement is possible at all times.
Indicator
4-digit LED: Status indicators for each channel,
4-zeiliges LCD: Datenspeicher auslesbar über RS 232
Communication
RS 232
1 analogue output per gas type, Ausgang 4–20 mA, linearisiert
data memory with history function
Operating temperature range
Ambient:
10/40 °C
Storage:
-25/+50 °C
Supply voltage
AC 230 V, 50 Hz
optional:
AC 115 V, 60 Hz
Power input:
max. 85 VA
Dimensions
W x H x D:
300 x 400 x 185 mm
Degree of protection
IP 54 (EN 60529)