CAPBs® set 4 Pa test

AFRISO CAPBs Set 4 Pa sens VOR 3190 22890 object_image_98542imagemain_en AFRISO CAPBs Set 4 Pa sens VOR 3190 22890 object_image_98542imagemain_en
CAPBs Set 4 Pa sens
AFRISO CAPBs Set 4 Pa sens VOR 3190 22890 object_image_98542imagemain_en
AFRISO CAPBs Set 4 Pa sens VOR 3190 22890 object_image_98542imagemain_en
Data sheet CAPBs® set 4 Pa test
AFRISO Data sheet CAPBs® set 4 Pa test

Versions

Part no. Remember
CAPBs® set 4 Pa sens P00119

Blue part no. = in-stock items

Description Format Language Download / Link
Data sheet CAPBs® set 4 Pa test Data sheet PDF
AFRISO Data sheet CAPBs® set 4 Pa test
If multiple heat systems using indoor intake air and ventilation systems are operated in a closed space, this can result in a hazardous low pressure. With the 4 Pa test, the underpressure limit value of 4 Pa/8 Pa can be checked and documented.
Dimensions (W x H x D)
Sensor module:
42 x 74 x 35 mm
Case:
260 x 63 x 155 mm
Weight
0.4 kg
Measuring range
Pressure:
-20/+20 mbar
Temperature:
-50/+600 °C
Pitot:
0.2/20 m/s
Accuracy
Pressure:
± 0.003 mbar (< 0.6 bar)
± 0.5 % v. Mw. ± 1 Digit (> 0.6 mbar)
Temperature:
± 1.5 K
Pitot:
± 0.1 m/s (< 2 m/s)
± 5 % v. Mw. (> 2 m/s)
Resolution
Pressure:
0.001 mbar
Temperature:
0.1 °C
Pitot:
0.1 m/s
Indication of measured values
hPa, mbar, Pa, mmWs, mmHg, inHg
Scope of delivery
  • Sensor module pressure/temperature CAPBs<sup>®</sup> sens FP 10 (20 mbar)
  • Hose kit SLS 4 Pa test
  • System case S