Skip to main content

RF100Si D402

Rohrfeder-Sicherheitsmanometer Typ D4

housing ⌀ (nominal size)100 mm
ConnectionG½B
alignmentbottom
accuracy class1,0
standardEN 837-1

  • Safety housing S3 as per EN 837-1/9.7.2
  • measuring system fully welded to housing
  • tightness-tested with helium
  • DNV- and GOSSTANDART-certified
  • Ex version (optional)

For corrosive gaseous and liquid media which are not highly viscous and do not crystallise; suitable for corrosive environments. This gauge is designed for applications as per EN 837-1/9.7.2.

RangeMounting typeTypePart number
-1/0 bardirectRF100Si D40285451402
-1/0.6 bardirectRF100Si D40285452402
-1/+1.5 bardirectRF100Si D40285453402
-1/+3 bardirectRF100Si D40285454402
-1/+5 bardirectRF100Si D40285455402
-1/+9 bardirectRF100Si D40285456402
-1/+15 bardirectRF100Si D40285457402
0/0.6 bardirectRF100Si D40285459402
0/1 bardirectRF100Si D40285460402
0/1.6 bardirectRF100Si D40285461402
0/2.5 bardirectRF100Si D40285462402
0/4 bardirectRF100Si D40285463402
0/6 bardirectRF100Si D40285464402
0/10 bardirectRF100Si D40285465402
0/16 bardirectRF100Si D40285466402
0/25 bardirectRF100Si D40285467402
0/40 bardirectRF100Si D40285468402
0/60 bardirectRF100Si D40285469402
0/100 bardirectRF100Si D40285470402
0/160 bardirectRF100Si D40285471402
0/250 bardirectRF100Si D40285472402
0/400 bardirectRF100Si D40285473402
0/600 bardirectRF100Si D40285474402
0/1 000 bardirectRF100Si D40285475402
  • in-stock items
  • manufactured goods

Type
D4
Nominal size
100 - 100
accuracy class
EN 837-1/6
NG 100: 1,0
ranges (EN 837-1/5)
see ordering table
application area
NG 100, NG 160
static load
≤ 600 bar:
Full scale value
> 600 bar:
¾ x full scale value
dynamic load
≤ 600 bar:
0.9 x full scale value
> 600 bar:
⅔ x full scale value
short-term
≤ 600 bar:
1.3 x full scale value
> 600 bar:
Full scale value
Operating temperature range
Medium:
max. 100 °C
Ambient:
-20/+60 °C
Degree of protection
IP 54 (EN 60529)

Standard design

Connection
Stainless steel 316 L, bottom
Measuring element
Bourdon tube
stainless steel 316 Ti/316 L
≤ 60 bar:
"C" type tube
> 60 bar:
helical tube
tightness-tested with helium
movement
Stainless steel
Dial
Aluminium, weiß
Scaling black
Pointer
Aluminium, schwarz
Housing
Edelstahl 304, with solid baffle wall and blow-out (S3)
bayonet bezel
Edelstahl 304
window
Sicherheitsverbundglas

Click the corresponding image for detail views, additional information and dimension tables.

AFRISO_SELECTION_PRESSURE_GAUGES_EN837-2_EN_11-16.PDF

Auswahlkriterien und Sicherheitsaspekte für Druckmessgeräte nach EN 837-2

PDF
854_017_0336-00.PDF

Rohrfeder-, Kapselfeder-, Plattenfeder-, Membranfedermanometer für Differenzdruck

PDF