Pressure transducers HydroFox® DMU 08

Afriso Pressure transducers HydroFox® DMU 08 Afriso Pressure transducers HydroFox® DMU 08
Afriso Legende: <div class="gallery-information"> <ol> <li>Junction box with pressure relief port</li> <li>Digital display and control unit VarioFox 12/14</li> <li>Signalling device</li> <li>HydroFox<sup>®</sup> DMU 08</li> </ol> </div>
Afriso Pressure transducers HydroFox® DMU 08 Afriso Pressure transducers HydroFox® DMU 08
Afriso Legende: <div class="gallery-information"> <p> <strong>Anchor clamp</strong> </p> <p>Dimensions (in mm)</p> <ol> <li>Cable diameter</li> </ol> </div>
Afriso Pressure transducers HydroFox® DMU 08 Afriso Pressure transducers HydroFox® DMU 08
Afriso Legende: <div class="gallery-information"> <p> <strong>Junction box with pressure relief port</strong> </p> <p>Dimensions (in mm)</p> <ol> <li>Earth</li> <li>Cable gland</li> <li>Pressure compensation element</li> </ol> </div>
Afriso Pressure transducers HydroFox® DMU 08
Afriso Pressure transducers HydroFox® DMU 08
Afriso Pressure transducers HydroFox® DMU 08
Afriso Pressure transducers HydroFox® DMU 08
Afriso Pressure transducers HydroFox® DMU 08
Afriso Pressure transducers HydroFox® DMU 08
Data sheet Pressure transducers HydroFox® DMU 08
Afriso Data sheet Pressure transducers HydroFox® DMU 08

Versions

Measuring range Overload Part no. Remember
HydroFox® DMU 08 with 5 m PUR cable 0/100 mbar At least 5 x FS 31555
HydroFox® DMU 08 with 5 m PUR cable 0/160 mbar At least 6 x FS 31556
HydroFox® DMU 08 with 5 m PUR cable 0/200 mbar At least 5 x FS 31557
HydroFox® DMU 08 with 5 m PUR cable 0/250 mbar At least 4 x FS 31558
HydroFox® DMU 08 with 5 m PUR cable 0/300 mbar At least 3 x FS 31519
HydroFox® DMU 08 with 5 m PUR cable 0/400 mbar At least 5 x FS 31559
HydroFox® DMU 08 with 5 m PUR cable 0/600 mbar At least 8 x FS 31560
HydroFox® DMU 08 with 5 m PUR cable 0/1 bar At least 5 x FS 31561
HydroFox® DMU 08 with 5 m PUR cable 0/1.6 bar At least 6 x FS 31562
HydroFox® DMU 08 with 5 m PUR cable 0/2 bar At least 4 x FS 31563
HydroFox® DMU 08 with 5 m PUR cable 0/2.5 bar At least 4 x FS 31564
HydroFox® DMU 08 with 5 m PUR cable 0/4 bar At least 5 x FS 31565
HydroFox® DMU 08 with 5 m PUR cable 0/6 bar At least 6 x FS 31566
HydroFox® DMU 08 with 5 m PUR cable 0/10 bar At least 4 x FS 31567
HydroFox® DMU 08 with 5 m PUR cable 0/16 bar At least 5 x FS 31568
HydroFox® DMU 08 with 5 m PUR cable 0/20 bar At least 3 x FS 31569
HydroFox® DMU 08 with 5 m PUR cable 0/25 bar At least 3 x FS 31570

Blue part no. = in-stock items

Description Format Language Download / Link
Data sheet Pressure transducers HydroFox® DMU 08 Data sheet PDF
Afriso Data sheet Pressure transducers HydroFox® DMU 08
Submersible probe DMU 08, DMU 08 T, DMU 09 Operating instructions PDF
Afriso Submersible probe DMU 08, DMU 08 T, DMU 09
Probes for is areas DMU 08…Ex, DMU 09…Ex Operating instructions PDF
Afriso Probes for is areas DMU 08…Ex, DMU 09…Ex
Electronic pressure measurement - pressure transducers DMU series Product range industrial technology PDF
Afriso Electronic pressure measurement - pressure transducers DMU series
Enquiry pressure transducers DMU PDF
Afriso Enquiry pressure transducers DMU
The level indicators HydroFox® DMU 08 use silicon technology and feature calibrated, amplified sensor signals which are available as standardised voltage or current outputs.
Measuring accuracy
Deviation from the characteristic curve according to IEC 60770 – limit point calibration (non-linearity, hysteresis, repeatability)
≤ 0/400 mbar < ± 0.5 % FSO
> 0/400 mbar < ± 0.35 % FSO
Measuring range
See ordering table
Operating temperature range
Medium:
-10/+70 °C
Ambient:
-10/+70 °C
Storage:
-25/+70 °C
Temperature error band
≤ 400 mbar:
≤ ± 1 % FSO/10 K In compensated range 0/70 °C
> 400 mbar:
≤ ± 0.75 % FSO/10 K In compensated range 0/70 °C
Dynamic characteristics
Response time:
≤ 10 ms
Probe
L x ⌀:
117 x 24 mm
Weight:
620 g
Material
Housing:
Stainless steel 316 L
Diaphragm:
Stainless steel 316 L
Seals:
FKM (Viton)
Cable:
PUR
Pressure transmission liquid
Silicone oil
Supply voltage
DC 12 – 36 V
Output signal
4–20 mA, 2-wire
Load
R<sub>Max</sub> = [(UB – UB<sub>Min</sub>)/0.02 A] Ω
Current input
< 25 mA
Electrical protection
Short circuit proof and protected against reverse polarity
Electrical connection (PUR cable with integrated breather tube for reference to the ambient atmospheric pressure)
Cable length:
5 m
Degree of protection:
IP 68 (EN 60529)
CE conformity
EMC Directive 2014/30/EU
RoHS Directive 2011/65/EU
Pressure Equipment Directive 2014/68/EU

Options

  • Extended weight
  • Measuring accuracy 0.1 % FSO
  • FEP cable
  • ATEX version
  • Screw connector kit
  • Junction box
  • Anchor clamp
Anchor clamp
Anchor clamp
Afriso
Dimensions (in mm)
  1. Cable diameter
Junction box with pressure relief port
Junction box with pressure relief port
Afriso
Dimensions (in mm)
  1. Earth
  2. Cable gland
  3. Pressure compensation element
DMU 08 - standard version
DMU 08 - standard version
Afriso
Dimensions (in mm)
Screw connector kit
Screw connector kit
Afriso
wiring diagram
wiring diagram
Afriso
2-wire, 4-20 mA